论文标题
一个自动化系统,以定义低温量热量器的最佳操作设置
An automated system to define the optimal operating settings of cryogenic calorimeters
论文作者
论文摘要
用NTD热敏电阻器械仪器已经开发了数十年的低温宏观钙化器。最佳偏置电流的选择对于能量分辨率和稳定性方面的适当操作至关重要。在本文中,我们提供了一组自动测量和分析程序,以表征和优化NTD热敏电阻的工作配置。提出的程序是针对Cuore开发的,Cuore是988个低温宏观钙化器,并使用NTD热敏电阻的仪器进行了仪器,这些热敏电阻自2017年以来一直在获取数据。这些程序使以可靠的方式表征了大量检测器。它们足够合适,还可以在其他大型低温探测器(例如丘比特)中使用。
Cryogenic macro-calorimeters instrumented with NTD thermistors have been developed for several decades. The choice of the optimal bias current is crucial for a proper operation of these detectors, both in terms of energy resolution and stability. In this paper we present a set of automatic measurements and analysis procedures for the characterization and optimization of the working configuration of the NTD thermistors. The presented procedures were developed for CUORE, an array of 988 cryogenic macro-calorimeters instrumented with NTD thermistors that has been taking data since 2017. These procedures made it possible to characterize a large number of detectors in a reliable way. They are suitable enough to be used also in other large arrays of cryogenic detectors, such as CUPID.