论文标题
闪烁探测器测量的意外巧合在宇宙射线上
Unexpected Coincidences in Scintillation Detector Measurements on Cosmic Rays
论文作者
论文摘要
我们使用两个位于各个位置的矩形闪烁检测器,测量了源自宇宙射线的低振幅信号。这些信号通过项目“ Muonlab”项目略微修改的信号分析仪进行了分析,该信号旨在衡量高中教育中宇宙辐射的兆次寿命和速度。在我们的实验中,我们集中在两个闪烁检测器中信号之间的时间差“三分法(DT)”和一个检测器中信号的信号振幅“脉冲高度(pH)”之间测量时间差的可能性。我们以两个排列进行了测量:第一个垂直,探测器平行,第二,第二水平,探测器平行,彼此相邻。我们观察到,在垂直和水平排列中,两个检测器中宇宙射线的低振幅信号都表现出意外的巧合。在正常的实验条件下,未观察到这些效应,因为有意选择光电倍增子(PMT)的扩增以及阈值电压,以避免噪声和过高的计数速率。巧合出现的速率非常小,比单个检测器的计数速率高一千倍。但是,三级频谱的测量似乎是检测这些罕见和意外影响的敏感方法。通过分辨率为0.5 ns测量两个分析仪输入的信号之间的时间差,脉冲高度测量的分辨率约为8 mV。据估计,巧合中涉及的低振幅信号必须响应闪烁体材料中100 keV的能量吸收。
We measured signals of low amplitudes originating from cosmic rays, using two rectangular-block scintillation detectors at various positions. The signals were analyzed by a slightly modified signal analyzer from project 'MuonLab', designed to measure the lifetime and velocity of muons from cosmic radiation in high school education. In our experiment we focused on the possibilities of the apparatus to measure the time difference 'deltatime (DT)' between signals in two scintillation detectors and the signal amplitude 'pulse height (PH)' of the signal in one detector. We performed measurements in two arrangements: first vertical, detectors parallel and above each other, second horizontal, detectors parallel and next to each other. We observed that in both the vertical and horizontal arrangements, low-amplitude signals from cosmic rays in the two detectors showed unexpected coincidences. Under normal experimental conditions these effects were not observed because the amplification in the photomultiplier (PMT) as well as the threshold voltage were intentionally chosen to avoid noise and excessively high count rates. The rate at which the coincidences occur is very small, more than a thousand times smaller than the count rates of the individual detectors. However, the measurement of the deltatime spectrum appears to be a sensitive way to detect the presence of these rare and unexpected effects. The time differences between the signals at the two analyzer inputs were measured with a resolution of 0.5 ns and the pulse height measurement had a resolution of approximately 8 mV. It is estimated that the low-amplitude signals involved in the coincidences must respond to energy absorptions of the order of 100 keV in the scintillator material.